Catalog Mitutoyo 2017 - 2019

Laser Scan Micrometer Measuring Unit

Series 544 LSM-503S

Specifications Linearity at 20°C *¹

Entire range : ±1 µm Narrow range : ±(0,6+0,1 Δ D) µm

High Accuracy Non-contact measuring system – General-purpose type with a measuring range of 0,3 mm to 30 mm. – Provides high accuracy with a linearity of ±1 µm over the entire measuring range and ±(0,6+0,1 Δ D) µm in the narrow range.

Positional error *² Repeatability (±2 ơ ) Measuring region Laser wavelength

±1,5 µm ±0,11 µm

10 x 30 mm

650 nm, Visible

– Excellent repeatability of ±0,1 µm. – High scanning rate of 3200 scans/s.

Scanning rate Scanning speed

3200 scans/s

226 m/s IEC, FDA

Applicable laser standards

Optional accessories No.

50

6

22

Description

18.5

10.5 15.5

Calibration Gauge Set, Ø 1 - 30 mm

02AGD130 02AGD490 02AGD440 02AGD450 02AGD240

68

42

53

202

57

Workstage for LSM

259

48

48

Centre support LSM-501S/503S/902

355

70min.

130

V-Block

350max.

80

170

55

50

Air Shield LSM-503S

8

8

10

42

LSM-503S

957608 Air Cleaner for Air Shield 02AGC150A Relay Extension Cable 02AGC150B Relay Extension Cable 02AGD440 and 02AGD450 : use with an adjustable workstage

30

80 5.5

90 20

110

12

57.5

95.5

84.5

46

65.5

54.5

250

105

Metric

No. Resolution (selectable) 544-536 0,3-30 0,00002-0,1 mm Range [mm]

Series 544 LSM-506S

Specifications Linearity at 20°C *¹

Entire Range : ±3 µm Narrow range : ±(1,5+0,5 Δ D) µm

High Accuracy Non-contact Measuring System – General-purpose type with a measuring range of 1 mm to 60 mm. – Provides high accuracy with a linearity of ±3 µm over the entire measuring range and ±(1,5+0,5 Δ D) µm in the narrow range.

Positional error *² Repeatability (±2 ơ ) Measuring region Laser wavelength

±4 µm

±0,36 µm

20 x 60 mm

650 nm, Visible

– Excellent repeatability of ±0,36 µm. – High scanning rate of 3200 scans/s.

Scanning rate Scanning speed

3200 scans/s

452 m/s IEC, FDA

Applicable laser standards

Optional accessories No.

10

50

M3 (x12)

Description

7

22

18.5

Calibration Gauge Set, Ø 1 - 60 mm

02AGD140 02AGD520 02AGD580 02AGD590 02AGD250

15.5 10.5 66 360 540

42

58

70

Workstage for LSM

258

102

Centre support LSM-506S/9506

90

90

V-Block

70min.

273 700max.

LSM-506S

Air Shield LSM-506S

185

110

163

82

957608 Air Cleaner for Air Shield 02AGC150A Relay Extension Cable 02AGC150B Relay Extension Cable 02AGD580 and 02AGD590 : use with an adjustable workstage

8

42

20

8

60

97

20 110

130

12

Base

7

115.5

168.5

71

87

87

*¹ At the centre of the measuring region *² Potential measurement error due to variation in workpiece position within the measuring region Δ D = Difference in diameter between the mastergauge and workpiece

348

192

Metric

No. Resolution (selectable) 544-538 0,00005-0,1 mm

Refer to the LSM brochure

385